Editor's Choice


Optical liquid analysis prototyping platform for ubiquitous sensing

30 March 2022 Editor's Choice Analogue, Mixed Signal, LSI

This drive toward ubiquitous, real-time sensing has shifted requirements for liquid sensors to smaller size, higher robustness and lower power, while still delivering high-quality results.

As the industry advances, intelligent platforms for on-the-go sensing are needed. These platforms need to be highly versatile, capable of satisfying unique requirements for a wide range of applications from environmental waters to process control. This article will introduce a portable, real-time sensing solution and prototyping platform for rapid liquid sensing.

A common liquid analysis technique

Various methods exist to test liquids, the purpose being to measure the concentration of an unknown parameter in a sample, such as pH, fluorescence or turbidity. A popular method is to evaluate liquids optically, as it is non-invasive and provides stable and accurate results.

Precision optical liquid measurements require mixed domain knowledge in electronics, optics and chemistry. In simple terms, the analysis begins with a sample that is exposed to light from a source such as an LED. After interacting with the sample, the resulting light is processed by a photodiode. That measured response is plotted against the measured responses of a set of standard samples of known concentrations. This is known as a calibration curve (Figure 1).

Using the calibration curve, the unknown value can be determined. This describes the general laboratory method for analytical measurements, but to meet the needs of ubiquitous sensing it must be scaled to different analytes and measurement techniques, as well as fit into a small form factor, all of which increases the complexity of design and evaluation.

Modular solution for rapid liquid measurement

The ADPD4101 from Analog Devices is an optical analog front end (AFE) capable of driving LEDs and synchronously receiving and processing signals from photodiodes to make high-precision optical measurements. The chip is highly configurable, featuring a high optical signal-to-noise ratio of up to 100 dB and high ambient light rejection provided by on-chip synchronous detection methods, allowing it to be used without an optically dark enclosure in many cases.


Figure 2. The CN0503 evaluation board.

The CN0503 reference design (Figure 2) was created to enable rapid prototyping of liquid analysis measurements with the ADPD4101. The design features the ADPD4101 as its core product, but adds up to four modular optical paths as well as measurement firmware and application software targeted for liquid analysis. The CN0503 interfaces directly with the ADICUP3029 board, which manages the measurement routine and data flow. The ADICUP3029 board can be connected directly to a laptop to view results in the evaluation GUI.

The CN0503 can measure fluorescence, turbidity, absorbance and colourimetry. A sample is prepared in a cuvette and placed in the 3D-printed cuvette holder, which houses the optics, including a lens and beam splitter. The cuvette holder slots into the appropriate optical path for plug-and-play measurement. In addition, the LED and photodiode cards can be switched out for even more customisation.

To demonstrate the creation of calibration curves and measurement of unknowns with the CN0503, measurements of pH, turbidity and fluorescence will be shown hereafter. The evaluation GUI was used to take measurements to create calibration curves. The noise value and limit of detection (LOD) were calculated to determine the lowest concentration feasible to be detected by the CN0503 for each example.

Measuring pH with absorbance

Absorbance background

Absorbance involves determining the concentration of a known solute in a solution based on how much light is absorbed at a particular wavelength. The concentration is proportional to the absorbance, per the Beer-Lambert Law. Many colourless analytes can be measured by adding a colour-changing reagent.

The example under examination in this article is measuring pH, one of the most common parameters measured across many industries, from water quality to wastewater treatment. Absorbance measurements are used for many other parameters, including dissolved oxygen/biological oxygen demand, nitrates, ammonia and chlorine.

Optics


Figure 3. An optical light path for measuring absorbance.

The optical path configuration for absorbance measurements is shown in Figure 3. With the CN0503, absorbance measurements can be made in any optical path (1 through 4). The incident beam is directed at the beam splitter, where a reference photodiode samples the intensity of the beam. The remaining power is directed through the sample. Taking the ratio of sample to reference light removes variation and noise of the LED source and synchronous pulse and receive windows provide ambient light rejection.


Figure 4. The CN0503 taking a pH measurement.

Set-up equipment:

• CN0503 evaluation board (Figure 4).

• EVAL-ADICUP3029 evaluation board.

• API pH test and adjuster kit.

• pH standards.

In this experiment, a colour indicator (bromothymol blue) was added to prepared solutions with different pH values. The solutions were transferred into cuvettes and tested at two different wavelengths – 430 nm and 615 nm

– where the indicator shows absorption changes vs. pH. The CN0503 makes this easy; the two different-wavelength LED cards can be inserted into optical path 2 and optical path 3, then the cuvette holder is simply moved into different paths for the different measurements.

Results

Using the CN0503 evaluation GUI, measurement results from both optical paths were easily exported into Excel. The resulting calibration curves for the two different wavelengths are shown in Figure 5 and Figure 6.

In each case, the pH vs. absorbance was plotted to create the calibration curve. The ‘Add Trendline’ option is then used to get an equation for the curve and the equations are then used to determine concentrations of unknown samples. The sensor output is the ‘x’ variable and the resulting ‘y’ value is the pH. This can be done by hand; however, the CN0503 can also be used for this. The firmware implements two fifth-order polynomials, INS1 and INS2. Once the polynomials are stored, the INS1 or INS2 mode can be selected so that measurement results are reported directly in the desired unit – in this case, pH. This makes it simple to quickly get a result for an unknown sample.

To get the noise value, two different data points were chosen for each wavelength: one lower pH value and one higher pH value. Two points were used because the curve fit was not linear in this case. The standard deviation of a set of repeated measurements for each point, reported as the noise value in Table 1, describes the precision of the measurement, excluding variations in sample preparation.

LOD is typically determined by measuring noise at low concentration and multiplying by three to give a confidence interval of 99,7%. Because pH is a logarithmic scale, pH 7 was chosen as the number to check the LOD, as shown in Table 2.


Table 2. pH measurement limit of detection.

Measuring turbidity

Turbidity background

Turbidity measurements of a liquid sample use the light scattering property of particles suspended in the liquid. Ultimately, it is a measure of the relative clarity of a liquid. The amount of light scattered and the scattering angle differ based on particle size, concentration and the wavelength of incident light. Measuring turbidity is conducted in many industries, including water quality and life sciences. Besides general turbidity, the CN0503 could be used to determine algae growth by measuring optical density.


Figure 7. Optical path for turbidity measurement.

Optics

Figure 7 shows the optical path for turbidity measurement with a 90° or 180° detector. With the CN0503, turbidity measurements can be made in optical path 1 or 4 only, as a 90° detector is needed. There are various measurement configurations and standards for turbidity. This example demonstrates a modified version of EPA Method 180.1, calibrated and reported in nephelometric turbidity units (NTU).


Figure 8. Turbidity calibration standards

Set-up equipment:

• CN0503 evaluation board.

• EVAL-ADICUP3029 evaluation board.

• Hanna Instruments’ turbidity standard calibration set (Figure 8).

For this experiment, optical path 4 was used with the 530 nm LED board inserted for testing.

Results

Using the CN0503 evaluation GUI, measurement results were exported into Excel. The resulting calibration curve is shown in Figure 9.

The response curve was split into two sections because the 90° scattering measurement is less responsive to high turbidities. One section represents lower turbidity (0 NTU to 100 NTU) and the other represents higher turbidity (100 NTU to 750 NTU). Two linear fits were then made to each section.


Table 3. Turbidity measurement noise value.

Even though there are now two equation values, the CN0503 can still be used to quickly show resulting NTU values. This is because each optical path can store its own equation values in INS1 and INS2. One important note is that INS1 and INS2 are dependent. The result of the first equation, INS1, is the input variable for the second equation, INS2. Once the equation values are stored, INS1 can be used to measure lower-turbidity samples and INS2 for higher-turbidity samples.

To get the noise value, we choose a data point to take the standard deviation of repeated measurements. The standard deviation is the noise value. One data point is chosen near the bottom of the range because the equation fit is linear.


Table 4. Turbidity measurement limit of detection

To determine the LOD, the noise value is measured for a blank or low concentration sample (Table 3), then multiplied by three to represent a 99,7% confidence interval (Table 4).

Measuring fluorescence with spinach solutions

Fluorescence background

When light is shined into a sample containing fluorescent molecules, the electrons move into a higher energy state and then lose some of that energy before emitting light at a longer wavelength. The fluorescence emission is chemically specific and can be used to identify the presence and number of specific molecules in a medium. In this example, fluorescing chlorophyll was demonstrated by using spinach leaves. Among many applications, fluorescence measurements are common in biological assays, dissolved oxygen, chemical oxygen demand and detecting if pasteurisation has been successful in milk.

Optics

The optical path configuration for fluorescence measurements is shown in Figure 10. With the CN0503, absorbance measurements can be made in optical paths 1 or 4 only, which is because of the 90° detector. Usually, a fluorescence detector is positioned at 90° from the incident light and a monochromatic or long-pass filter is used to increase isolation between the excitation and emitted light. Fluorescence is a very sensitive low-level measurement and is subject to interference, so the reference detector and the synchronous detection methods are used to reduce error sources.

Set-up equipment:

• CN0503 evaluation board.

• EVAL-ADICUP3029 evaluation board.

• Spinach solution.


Figure 11. Chlorophyll samples made with spinach.

For this experiment, a spinach solution was created by blending spinach leaves with water. This was then filtered and kept as the stock solution. The stock solution was then diluted to result in samples of different percentages of spinach solution (Figure 11). These were used as standards to create a curve of % spinach solution through fluorescence. Optical path 1 was used along with the 365 nm LED card and the long-pass filter inserted.

Results

The calibration curve for % spinach solution is shown in Figure 12. The trendline equation for this calibration curve can be stored so that the results are directly reported as a percentage by the CN0503. To get the noise value, two different data points were picked: one near the bottom of the range and one near the top, because the curve fit was not linear. The noise is given by the standard deviation of a set of repeated measurements for each point, as shown in Table 5.

To determine the LOD, the noise value is measured for a blank or low concentration sample, then multiplied by three to represent a 99,7% confidence interval (Table 6).


Table 6. Fluorescence measurement limit of detection.

Conclusion

Prototyping complex optical liquid analysis measurements is a challenge that requires careful consideration of how chemistry, optics and electronics interact to produce a precise result. Integrated analog front-end products like the ADPD4101 pave the way for higher-performance optical liquid sensing in smaller spaces.

The CN0503 builds on the ADPD4101, including optical design, firmware and software for an easy-to-use and highly customisable rapid prototyping platform that is capable of taking accurate optical measurements of liquid parameters, including absorbance, colourimetry, turbidity and fluorescence.


Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Ultra-low-power wireless module
Altron Arrow Telecoms, Datacoms, Wireless, IoT
The STM32WBA5MMG from STMicroelectronics is an ultra-low-power, small form factor, certified 2,4 GHz wireless module that supports Bluetooth LE, Zigbee 3.0, OpenThread, and IEEE 802.15.4 proprietary protocols.

Read more...
16-channel multicell battery monitor
Altron Arrow Power Electronics / Power Management
The ADBMS6830B is a multicell battery stack monitor that measures up to 16 series-connected battery cells with a lifetime total measurement error of less than 2 mV.

Read more...
Enhance SiC device efficiency using merged-pin Schottky diodes
NuVision Electronics Editor's Choice Power Electronics / Power Management
Silicon carbide (SiC) has advantages over silicon (Si) that make it particularly suitable for Schottky diodes in applications such as fast battery chargers, photovoltaic (PV) battery converters, and traction inverters.

Read more...
Bringing Bluetooth Channel Sounding to automotive and beyond with KW47
Altron Arrow Telecoms, Datacoms, Wireless, IoT
NXP’s new Channel Sounding-certified KW47 and MCX W72 wireless MCUs are set to help automakers with distance measurement, bringing an additional ranging solution for car access and autonomous systems, and will be utilised across a broader spectrum of applications.

Read more...
What is Wi-Fi HaLow and why choose it for IoT?
iCorp Technologies Editor's Choice Telecoms, Datacoms, Wireless, IoT
Wi-Fi HaLow introduces a low power connectivity option that, in contrast to other Wi-Fi options, offers greater range of approximately 1 km, which opens up a raft of IoT use cases.

Read more...
Wi-Fi 6 and Bluetooth LE coprocessor module
Altron Arrow Telecoms, Datacoms, Wireless, IoT
The ST67W611M1 from STMicroelectronics boasts an all-in-one design which, together with its capabilities, contribute to making it an attractive choice for IoT edge devices requiring a single-chip solution.

Read more...
Wi-Fi 6 plus Bluetooth LE SoC
Altron Arrow Telecoms, Datacoms, Wireless, IoT
Silicon Labs’ SiWx917M SoC is the company’s lowest power Wi-Fi 6 SoC, ideal for ultra-low power IoT wireless devices using Wi-Fi, Bluetooth, Matter, and IP networking for secure cloud connectivity.

Read more...
Simple battery charger ICs for any chemistry
Altron Arrow Editor's Choice Power Electronics / Power Management
The LTC4162 is a highly integrated, high voltage multi-chemistry synchronous monolithic step-down battery charger and PowerPath manager with onboard telemetry functions and optional maximum power point tracking.

Read more...
Microchip enhances TrustMANAGER platform
Altron Arrow DSP, Micros & Memory
Firmware over-the-air updates and remote cryptographic key management provide scalable solutions for addressing IoT security challenges.

Read more...
Adaptive optics’ power solution
Altron Arrow Opto-Electronics
Vicor power-dense adaptive optical modules enable colossal telescopes to look into the past for deep space discoveries.

Read more...









While every effort has been made to ensure the accuracy of the information contained herein, the publisher and its agents cannot be held responsible for any errors contained, or any loss incurred as a result. Articles published do not necessarily reflect the views of the publishers. The editor reserves the right to alter or cut copy. Articles submitted are deemed to have been cleared for publication. Advertisements and company contact details are published as provided by the advertiser. Technews Publishing (Pty) Ltd cannot be held responsible for the accuracy or veracity of supplied material.




© Technews Publishing (Pty) Ltd | All Rights Reserved